Agilent Technologies 54621A, 54624A, 54621D, 54622D General Information, Analog Channel Triggering

Models: 54621A 54622A 54622D 54621D 54624A

1 111
Download 111 pages 42.09 Kb
Page 19
Image 19
Analog Channel Triggering

General Information

Analog Channel Triggering

*Denotes Warranted Specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ±10 °C from firmware calibration temperature.

Analog Channel Triggering

Range (Internal)

±6 div from center screen

Sensitivity*

Greater of 0.35 div or 2.5 mV

Coupling

AC (~3.5 Hz), DC, noise reject, HF reject and LF reject (~ 50 kHz)

Digital (D15 - D0) Channel Triggering (54621D and 5462 2D)

Threshold Range (user-defined)

±8.0 V in 10 mV increments

Threshold Accuracy*

±(100 mV + 3% of threshold setting)

Predefined Thresholds

TTL = 1.4 V, CMOS = 2.5 V, ECL = -1.3 V

External (EXT) Triggering

Input Resistance

1 M±3%

Input Impedance

~ 14 pF

Maximum Input

CAT I 300 Vrms, 400 Vpk

 

CAT II 100 Vrms, 400 Vpk

 

with 10074C 10:1 probe:CAT I 500 Vpk, CAT II 400 Vpk

Range

±10 V

Sensitivity

dc to 25 MHz, < 75 mV

 

25 MHz to max bandwidth, < 150 mV

Coupling

AC (~ 3.5 Hz), DC, noise reject, HF reject and LF reject (~ 50 kHz)

Probe ID (Agilent/HP & Tek

Auto probe sense for 54621A/22A

Compatible)

 

1-14