Agilent Technologies E1439 manual Option/fiberMode Off, Append

Models: E1439

1 246
Download 246 pages 11.5 Kb
Page 110
Image 110

Agilent E1439 Programmer's Reference

Functions listed alphabetically

headerValuePtr points to the current value of initalValue

incrementCount specifies the number of automatic increments to the headerIndex bit field. The default incrementCount is 0 and the maximum value is (2^10 -1).

Example

The following is a example of how the increment process works.

For headerValue = 0x12345678 and incrementCount = 0x2, the sequence of values for headerValue and headerIndex are:

Increment

headerValue

headerIndex

0

0x12345678

0x278

1

0x12345679

0x279

2

0x1234567A

0x27A

0

0x12345678

0x278

1

0x12345679

0x279

2

0x1234567A

0x27A

 

If an incremented header reaches a value where the headerIndex is 0x3FF, the next headerIndex

 

will be 0x000, and no carry will be generated to the upper 22 bits of the header.

Note

If the incrementCount is set to 0, incrementing the headerIndex field is disabled.

incrementCountPtr points to the current value of incrementCount

The following table is a summary of valid fiber, epoch setups. Please note that the designation of N/A means that this information is not applicable to this condition. In this case the setting is accepted but ignored. The designation of OK means the setting is accepted and implemented. The designation of NO means do not use this setting with this condition.

Option/fiberMode

Off

Copy1

Raw

Generate

Append

BOF_OFF1

N/A

N/A

OK

OK

OK

BOF_ON2

N/A

N/A

OK

OK3

OK4

CRC_OFF

N/A

OK5

OK

OK

OK5

CRC_ON1

N/A

OK5

OK

OK

OK5

FLOW_CONTROL_OFF1

N/A

N/A

OK

OK

OK

FLOW_CONTROL_COPY

N/A

N/A

OK

OK

OK

FLOW_CONTROL_NO_COPY

N/A

N/A

OK

OK

OK

EPOCH_GEN_OFF

N/A

N/A

OK

OK

NO

EPOCH_GEN_ON1

N/A

N/A

OK

OK3

OK6

HEADER_OFF1

N/A

N/A

OK

OK

OK

HEADER_ON2

N/A

N/A

OK

OK

OK

1.Default instrument setting on power-up and reset.

2.Not applicable unless EPOCH _GEN_ON is enabled.

100

Page 110
Image 110
Agilent Technologies E1439 manual Option/fiberMode Off, Append