Agilent Technologies E1439 manual Data type, Call age1439measstart Call age1439readraw

Models: E1439

1 246
Download 246 pages 11.5 Kb
Page 173
Image 173

 

Agilent E1439 Programmer's Reference

 

 

Functions listed alphabetically

Data type

Resolution

Words per sample

(bits)

 

 

 

 

 

 

 

REAL

12

2

 

REAL

24

4

 

COMPLEX

12

4

 

COMPLEX

24

8

 

In continuous data collection mode, wordCount should be set equal to the maximum possible wordCount to ensure that the entire data block is read out.

overloadPtr

returns an overload indicator. See “Comments on Overload” on page 160. The way to properly

 

use the overload argument for the age1439_read_raw function is this:

1. Set up the hardware.

2. Call age1439_meas_start.

3. Call age1439_read_raw.

If data is not available, the read function returns immediately with one of the following return values, and the overload indication is AGE1439_OFF:

AGE1439_NO_DATA_MEASUREMENT_IN_PROGRESS

AGE1439_NO_DATA_MEASUREMENT_PAUSED

AGE1439_NO_DATA_WAITING_FOR_TRIGGER

AGE1439_NO_DATA_WAITING_FOR_ARM

When data is available, AGE1439_SUCCESS is returned and the overload value reflects whether an overload was encountered for the given data block.

4. In continuous mode, subsequent data blocks can be read by calling the

 

age1439_read_raw function again (age1439_meas_start should not be called

 

again).

 

5. When used in this way, an overload indication is given for each and every

 

data block read in block mode. In continuous mode the overload indicator

 

only means there was an overload sometime after calling age1439_meas_start.

 

 

Note

The primary purpose of the age1439_read_raw function is to provide the fastest possible way to

 

read blocks of data from the module. Since this command does not perform data scaling after

 

reading data it may save 10-20% of the overall age1439_read time, depending on the host

 

computer in use. The resulting data ordering is dependent on the data type and resolution. The

 

array may be cast as a long before reading the data to provide whole words.

 

 

Example

Adeclaration in the Front Panel example program can be changed to exercise age1439_read_ raw() in frmMain of e1439.vbp:

Const constFreqCentRaw = False ’when TRUE, use age1439_read_raw()

’instead of age1439_read

Return Value

AGE1439_SUCCESS

AGE1439_NO_DATA_MEASUREMENT_IN_PROGRESS

163