Agilent E1439 Programmer's Reference

Functions listed alphabetically

age1439_clock_setup

Sets all timing parameters for commonly used measurement setups. This description also includes a query:

age1439_clock_setup_get

VXIplug&play Syntax

#include "age1439".h

ViStatus age1439_clock_setup(ViSession id, ViInt16 clockSetup); ViStatus age1439_clock_setup_get(ViSession id, ViPInt16 clockSetupPtr);

Description

age1439_clock_setup is used to select the source and distribution of clocking and synchronization signals used by the Agilent E1439 module. The primary clock signal used by the module is the ADC clock, for which the rising edges indicate the time for each sample of the analog-to-digital converter.

This function changes the settings controlled by the following lower-level functions:

age1439_adc_clock age1439_adc_divider age1439_front_panel_clock_input age1439_reference_clock age1439_reference_prescaler age1439_smb_clock_output age1439_sync_clock age1439_sync_direction age1439_sync_output age1439_vcxo

Setups using the external sample clock require that you use age1439_clock_fs to supply the clock frequency.

Parameters

is the VXI instrument session pointer returned by the age1439_init function.

This parameter provides a quick way to set up most of the timing parameters for several standard clock configurations. The following setups are available: