Agilent Technologies manual Status Bit Definition Description, AGE1439STATUSREADVALID

Models: E1439

1 246
Download 246 pages 11.5 Kb
Page 187
Image 187

 

 

Agilent E1439 Programmer's Reference

 

 

Functions listed alphabetically

Status Bit

Definition

Description

 

 

 

 

8

AGE1439_STATUS_READ_VALID

This flag is set whenever there is at least one valid 16-bit

 

 

data word available to be read via the VME data register. Not

 

 

valid when using the local bus data port.

9

AGE1439_STATUS_BLOCK_READY

This bit is set in continuous mode whenever the size of the

 

 

data in the FIFO is equal to or greater than the block size

 

 

register. Check this bit before reading data to insure that a

 

 

block of data may be transferred without fear of running out

 

 

of data, thereby holding up the Local bus or VME bus. This bit

 

 

is set in block mode whenever the module has successfully

 

 

taken a block size number of samples since the most recent

 

 

trigger and is cleared when the block is read out, when force

 

 

to Idle is asserted, or when the module is armed for another

 

 

measurement.

10

AGE1439_STATUS_ARMED

This bit is set whenever the module is in the Trigger state, or

 

 

is in the Arm state and has satisfied its pre-trigger

 

 

requirements. When this bit is set, the module releases the

 

 

VXI Sync line. Once all modules release the Sync line, then all

 

 

modules go to the Trigger state.

11

AGE1439_STATUS_FIFO_OVERFLOW

FIFO Overflow: This bit set when the FIFO buffer overflows

 

 

in continuous mode

12

AGE1439_STATUS_OVERLOAD

This bit is set whenever the ADC converts a sample that

 

 

exceeds the range of the ADC. The bit is cleared when the

 

 

Status register is read.

13

AGE1439_STATUS_ERROR_QUEUE

This bit is set whenever there is an error in the error queue. It

 

 

is cleared when the error queue is empty

14

AGE1439_STATUS_MODID

A (1) in this field indicates that the module is not selected via

 

 

the P2 MODID line. A (0) indicates that the module is

 

 

selected by a high state on the P2 MODID line

15

AGE1439_STATUS_HARDWARE_SET

This bit is set when all commands are complete and the

 

 

hardware has been set

Return Value

AGE1439_SUCCESS indicates that a function was successful.

Values other than AGE1439_SUCCESS indicate an error condition or other important status condition. To determine the error message, pass the return value to “age1439_error_message” on page 102.

See Also

“age1439_init” on page 132

177

Page 187
Image 187
Agilent Technologies manual Status Bit Definition Description, AGE1439STATUSREADVALID