Using the Agilent E1439

Managing multiple modules

Managing multiple modules

Sharing Reference and Sync signals in multi-module systems

The Agilent E1439 supports synchronous operation among multiple E1439s by using a shared ADC clock and Sync signal to drive all the modules in a system. The shared Sync signal is used to synchronize critical operations including arming, triggering the beginning of data collection, setting a common phase of the local oscillators for zoom operation, and forcing concurrent output sample times when decimation is used. The Sync line transitions are constrained to not occur during the critical (setup and hold) regions of the external reference. The reference operates at 1/38 of the internal ADC clock, typically 95 MHz for a E1439 module. The reference can be either generated within the master module or an external reference can be fed into the master module through a front panel BNC.

For optimal phase noise performance in multi-module systems it is recommended that the first channel be an Agilent E1439C or D1. The Agilent E1439C does not support local bus or fiber optic transfers.

Multi-module systems may include multiple Agilent E1438s or Agilent E1439s but not a mixture of the two types of modules.

Clock distribution

When shared, the reference clock and sync lines are distributed among modules either on the VXI backplane using the ECL Trigger lines, or on the front panel using the SMB Clock/Ref extender connectors. When VXI backplane distribution is used with more than one VXI mainframe, the front panel Intermodule Clock and Sync connectors can be used to distribute clock and Sync lines from one mainframe to another.

Since the Sync transition timing relative to the reference input is critical, the module driving the Sync line should ideally be the same one identified as the master. However, when using backplane distribution, any E1439 in the same mainframe as the master can drive the Sync line.

When using the multi-sync mode of operation, the selection of front panel or backplane distribution of reference and Sync signals involves the following considerations:

Backplane distribution requires the use of the ECL Trigger lines on the backplane, which are then unavailable to other modules.

The overall time skew between the arrival of ADC clock edges is smaller when using backplane distribution, particularly if the master (or buffer) module is physically located in the center of the group of E1439 modules.

Backplane distribution is more susceptible to pickup of jitter on the ADC clock from other digital activity on the VXI backplane. The extent of this pickup depends on the mainframe and on the other modules in the mainframe. One important step in reducing this pickup is to disable, whenever possible, the 10 MHz VXI clock generated by the slot-0 controller.