Using the Agilent E1439

Fiber Optic Interface

The following shows two E1439D modules in an append chain transmitting data to a fiber receiver when flow control is off.

 

E1438D / E1439D

Fiber RX

 

Fiber TX

 

 

1 KB

 

 

FIFO

ADC

DIGITAL

FIFO

FILTERS

 

VME BUS

 

 

 

 

LBUS

 

E1438D / E1439D

 

Fiber RX

 

1 KB

Fiber TX

 

FIFO

 

 

 

 

 

 

1 KB

 

 

 

FIFO

ADC

DIGITAL

FIFO

 

FILTERS

 

 

 

VME BUS

 

 

 

 

 

 

LBUS

Fiber Receiver

Fiber

Fiber

RX

TX

 

DATA

 

Processing

Fiber Interface Setup

 

 

First E1439D in chain

Next E1439D in chain

Mode

Generate

Mode

Append

Rate

106 or 250 MBs

Rate

same as module to left

BOF

ON

BOF

ON1

CRC

ON

CRC

ON

Flow Control

OFF

Flow Control

OFF

Epoch Generate

ON

Epoch Generate

ON

Epoch Size

Divisible by 4; must match blocksize

Epoch Size

Divisible by 4; must match blocksize

1.The final module in an append chain may require BOF to be Off for compat- ibility with data receivers that cannot process BOFs.