Agilent E1439 Programmer's Reference

Functions listed alphabetically

Variable

Range Index

Full Scale

Full Scale

(dBm)

Voltage (Vp)

 

 

 

 

 

 

AGE1439_RANGE_21

21

−15

.0562

AGE1439_RANGE_20

20

−16

.0501

AGE1439_RANGE_19

19

−17

.0447

AGE1439_RANGE_18

18

−18

.0398

AGE1439_RANGE_17

17

−19

.0355

AGE1439_RANGE_16

16

−20

.0316

AGE1439_RANGE_15

15

−21

.0282

AGE1439_RANGE_14

14

−22

.0251

AGE1439_RANGE_13

13

−23

.0224

AGE1439_RANGE_12

12

−24

.02

AGE1439_RANGE_11

11

−25

.0178

AGE1439_RANGE_10

10

−26

.0158

AGE1439_RANGE_9

9

−27

.0141

AGE1439_RANGE_8

8

−28

.0126

AGE1439_RANGE_7

7

−29

.0112

AGE1439_RANGE_6

6

−30

.01

AGE1439_RANGE_5

5

−31

.0089

AGE1439_RANGE_4

4

−32

.0079

AGE1439_RANGE_3

3

−33

.0071

AGE1439_RANGE_2

2

−34

.0063

AGE1439_RANGE_1

1

−35

.0056

AGE1439_RANGE_0

0

−36

.005

AGE1439_RANGE_MIN

0

−36

.005

Note

These values are approximate. For more accuracy use age1439_data_scale_get.

 

 

rangePtr

points to the current value of the range parameter for the selected signalPath. For the AGE1439_

 

BB_PATH signalPath the returned range is always AGE1439_RANGE_15.

signal

determines whether or not the input signal is connected to the input amplifier.

 

AGE1439_SIGNAL_ON attaches the input signal to the 50 Ohm buffer amplifier.

 

AGE1439_SIGNAL_OFF redirects the input signal to a dummy 50 Ohm load, and feeds the

 

buffer amplifier from an internally grounded 50 Ohm source resistance. The signal OFF setting is

 

useful for making reference measurements without the signal applied. When using ac coupling the

 

0.2 µF capacitor remains between the input connector and its 50 Ohm termination.

signalPtr

points to the current value of the signal parameter.

signalPath

Selects baseband AGE1439_BB_PATH or IF signal path AGE1439_IF_PATH. The IF path passes

 

frequencies between 52 and 88 MHz. The range values above only apply to the IF signal path.

signalPathPtr

points to the current value of signalPath