Installing the Agilent E1439

To install the Agilent E1439

To install the Agilent E1439

To protect circuits from static discharge, observe anti-static techniques whenever handling the Agilent E1439 VXI ADC Module.

1.Set up your VXI mainframe. See the installation guide for your mainframe.

2.Select a slot in the VXI mainframe for the E1439 module.

The Agilent E1439D module’s local bus receives ECL-level data from the module immediately to its left and outputs ECL-level data to the module immediately to its right. Every module using the local bus is keyed to prevent two modules from fitting next to each other unless they are compatible. If you will be using the local bus, select adjacent slots immediately to the left of the data-receiving module. If the VXI bus is used, maximum data rates will be reduced but the module can be placed in any available slot.

3.Using a small screwdriver or similar tool, set the logical address configura- tion switch on the E1439. (See the illustration on the next page.) Each module in the system must have a unique logical address. The factory default setting is 1100 0000 (192).

For optimal phase noise performance in multi-module systems it is recommended that the first channel be an Agilent E1439C or D1. The Agilent E1439C does not support local bus or fiber optic transfers.

Multi-module systems may include multiple Agilent E1438s or Agilent E1439s but not a mixture of the two types of modules.

1As opposed to the older A or B models.