Agilent E1439 Programmer's Reference

Functions listed alphabetically

age1439_error_message

Returns error information obtained from function calls.

VXIplug&play Syntax

#include "age1439".h

ViStatus age1439_error_message(ViSession id, ViStatus statusCode, ViChar

 

errorMessage[]);

 

Description

 

age1439_error_message takes an error return value generated by a function and translates it to a

 

readable string. This function includes host errors as well as firmware errors.

 

Parameters

id

is the VXI instrument session pointer returned by the age1439_init function.

errorMessage

represents the error message string up to 256 characters long.

 

 

Note

For this parameter you must allocate a character array of at least 256 characters AGE1439_STR_

 

LEN_MIN, including the null byte, prior to calling this function in any programming language.

 

 

statusCode

represents the instrument numeric error code.

 

Return Value

 

AGE1439_SUCCESS indicates that a function was successful.

 

Values other than AGE1439_SUCCESS indicate an unknown error condition or other important

 

status condition and may return VI_WARN_UNKNOWN_STATUS.

 

See Also

 

“age1439_init” on page 132, “age1439_error_query” on page 103, “Error messages” on page 199