id

refClock

refClockPtr

Agilent E1439 Programmer's Reference

Functions listed alphabetically

age1439_reference_clock

Selects the source of the reference clock. This description also includes the query function:

age1439_reference_clock_get

VXIplug&play Syntax

#include "age1439".h

ViStatus age1439_reference_clock(ViSession id, ViInt16 refClock); ViStatus age1439_reference_clock_get(ViSession id, ViPInt16 refClockPtr);

This command should be used only for specialized custom clock requirements. Most useful clock setups can be supplied by age1439_clock_setup.

Parameters

is the VXI instrument session pointer returned by the age1439_init function.

AGE1439_FRONT_PANEL_CLOCK specifies the front panel clock be used as the reference clock. Use this in conjunction with age1439_front_panel_clock_input.

AGE1439_VXI_CLOCK specifies that the VXI (rear panel) clock be used as the reference clock. Use this in conjunction with age1439_vxi_clock_output.

Returns a pointer to the current value of refClock.

Return Value

AGE1439_SUCCESS indicates that a function was successful.

Values other than AGE1439_SUCCESS indicate an error condition or other important status condition. To determine the error message, pass the return value to “age1439_error_message” on page 102.

See Also

“Default values” on page 201, “age1439_init” on page 132, “age1439_clock_setup” on page 78, “age1439_front_panel_clock_input” on page 131, “age1439_vxi_clock_output” on page 188, “age1439_reference_prescaler” on page 166, “Using clock and sync” in chapter 3