Signal

Amplifier 12 dB

Highpass

Filter

52 MHz

 

IF

 

Baseband

 

Lowpass

Bypass

Filter

!

0-36 MHz

Attenuator 1-16 dB

Amplifier

6dB

IF

Baseband

Module Description

Block diagram and description

Bandpass

Filter

52-88 MHz

Attenuator

1-16 dB

Lowpass

 

Filter

95 MHz

0-43 MHz

 

Sampling

 

ADC

 

Offset

DAC

AC DC

To Digital Filters

Clock Generation

The source for a clock signal is the 95 MHz crystal oscillator inside the E1439. This oscillator can free run or be locked to an external reference signal through the front-panel BNC "Ext Clock/Ref". This signal can be TTL, ECL, or sine wave. The oscillator can also be locked to a reference routed via the backplane. A 2.5 MHz reference signal is available to be routed out the front panel or the backplane to lock additional E1439s.