Agilent E1439 Programmer's Reference

Functions listed alphabetically

AGE1439_ADC inserts a 0.2 µF capacitor between the input connector and the 50 Ohm buffer amplifier.

couplingPtr

points to the current value of the coupling parameter for an Agilent E1439 or group of

 

Agilent E1439s.

 

 

 

 

range

is a range index number which is transformed to a full scale voltage value. This function always

 

sets only the IF signal path range even if signalPath is set to AGE1439_BB_PATH. In baseband

 

mode the range is fixed at -21 dBm.

 

 

 

 

 

AGE1438_RANGE_MAX sets the range to the maximum allowable.

 

 

 

AGE1439_RANGE_MIN sets the range to the minimum allowable.

 

 

 

Signal inputs with an absolute value larger than full scale generate an ADC overflow error.

 

Range values are as follows.

 

 

 

 

 

 

Variable

Range Index

Full Scale

Full Scale

 

 

(dBm)

Voltage (Vp)

 

 

 

 

 

 

 

 

 

 

 

 

 

AGE1439_RANGE_MAX

48

 

 

 

 

 

AGE1439_RANGE_48

48

12

1.26

 

 

 

AGE1439_RANGE_47

47

11

1.12

 

 

 

AGE1439_RANGE_46

46

10

1

 

 

 

AGE1439_RANGE_45

45

9

.891

 

 

 

AGE1439_RANGE_44

44

8

.794

 

 

 

AGE1439_RANGE_43

43

7

.708

 

 

 

AGE1439_RANGE_42

42

6

.631

 

 

 

AGE1439_RANGE_41

41

5

.562

 

 

 

AGE1439_RANGE_40

40

4

.501

 

 

 

AGE1439_RANGE_39

39

3

.447

 

 

 

AGE1439_RANGE_38

38

2

.398

 

 

 

AGE1439_RANGE_37

37

1

.355

 

 

 

AGE1439_RANGE_36

36

0

.316

 

 

 

AGE1439_RANGE_35

35

−1

.282

 

 

 

AGE1439_RANGE_34

34

−2

.251

 

 

 

AGE1439_RANGE_33

33

−3

.224

 

 

 

AGE1439_RANGE_32

32

−4

.12

 

 

 

AGE1439_RANGE_31

31

−5

.178

 

 

 

AGE1439_RANGE_30

30

−6

.159

 

 

 

AGE1439_RANGE_29

29

−7

.141

 

 

 

AGE1439_RANGE_28

28

−8

.126

 

 

 

AGE1439_RANGE_27

27

−9

.112

 

 

 

AGE1439_RANGE_26

26

−10

.1

 

 

 

AGE1439_RANGE_25

25

−11

.089

 

 

 

AGE1439_RANGE_24

24

−12

.0794

 

 

 

AGE1439_RANGE_23

23

−13

.0708

 

 

 

AGE1439_RANGE_22

22

−14

.0631

 

142

Page 152
Image 152
Agilent Technologies E1439 manual CouplingPtr, Range