Agilent E1439 Programmer's Reference

Functions listed alphabetically

age1439_combo_setup

Combines often used setup commands from various functions.

 

age1439_combo_setup sets signal bandwidth, blocksize and center frequency.

 

VXIplug&play Syntax

 

#include "age1439".h

 

ViStatus age1439_combo_setup(ViSession id, ViInt16 sigBw, ViInt32 blocksize, ViInt32

 

phase, ViInt32 interpolate);

 

Description

 

age1439_combo_setup provides a faster way to set up parameters from several functions which

 

are often used together.

 

Parameters

id

is the VXI instrument session pointer returned by the age1439_init function.

blocksize

See “age1439_data_setup” on page 90 for a description of the blocksize parameter.

interpolate

See “age1439_frequency_center_raw” on page 125 for a description of the interpolate parameter.

phase

See “age1439_frequency_center_raw” on page 125 for a description of the phase parameter.

sigBw

See “age1439_filter_setup” on page 120 for a description of the sigBw parameter.

 

Comments

 

This command halts the current measurement which also releases the forced Idle state. If you use

 

this command in multi-module systems to synchronously change the center frequency while the

 

modules are forced to Idle, then you should subsquently call age1439_meas_control to re-assert

 

the forced Idle condition.

 

Return Value

 

AGE1439_SUCCESS indicates that a function was successful.

 

Values other than AGE1439_SUCCESS indicate an error condition or other important status

 

condition. To determine the error message, pass the return value to “age1439_error_message” on

 

page 102.

 

See Also

 

“Commands which halt active measurements” on page 198, “age1439_init” on page 132,

 

“age1439_filter_setup” on page 120, “age1439_frequency_center_raw” on page 125, “age1439_

 

data_setup” on page 90, “age1439_meas_control” on page 151