Agilent E1439 Programmer's Reference

Functions listed alphabetically

age1439_status_get

Reads status register information for the module.

 

VXIplug&play Syntax

 

 

 

#include "age1439".h

 

 

 

ViStatus age1439_status_get(ViSession id, ViPInt16 statusPtr);

 

 

Parameters

 

 

id

is the VXI instrument session pointer returned by the age1439_init function.

statusPtr

points to the status word. The bits are defined below:

 

 

Status Bit

Definition

Description

0-1AGE1439_STATUS_MEAS_ARM_WAIT AGE1439_STATUS_MEAS_IDLE AGE1439_STATUS_MEAS_IN_

PROGRESS

AGE1439_STATUS_MEAS_TRIG_WAIT

These two bits indicate the current state of the measurement loop. See “The measurement loop” in chapter 3 and “age1439_meas_status_get” on page 156 for more information about these states

2

AGE1439_STATUS_PASSED

Passed: This bit is always set to 1

3

AGE1439_STATUS_READY

This bit is set when the module is ready after power-on. See

 

 

the VXIbus Specifications for more information.

4

AGE1439_STATUS_FIBER_ACTIVE

This bit is set internally whenever any data has been written

 

 

to the receive FIFO, or read from the transmit FIFO of the

 

 

fiber interface within the past 500 milliseconds,

 

 

approximately. The bit is cleared automatically when activity

 

 

ceases on the fiber interface

5

AGE1439_STATUS_FIBER_ERROR

This bit is set internally whenever an error condition occurs

 

 

on the fiber interface. Reading the status register does not

 

 

clear this bit. To do this, the age1439_fiber_error_clear

 

 

function must be used explicitly. The function age1439_

 

 

fiber_error_get can be used to read the contents of the fiber

 

 

error register. If the error is continuously present, the bit will

 

 

not be cleared.

6

AGE1439_STATUS_SETUP_ERROR

Setup error: An invalid parameter value was requested. If an

 

 

invalid block size was requested, the closest valid block size

 

 

is used until a change to an interrelated parameter makes the

 

 

requested block size valid. If a data resolution, data type,

 

 

filter bandwidth, trigger delay, or filter decimation parameter

 

 

was requested which would result in an inability to make a

 

 

measurement, the previous valid parameter is used until a

 

 

change to an interrelated parameter makes the requested

 

 

parameter valid

7

AGE1439_STATUS_SYNC_COMPLETE

Sync/Idle Complete: This bit is set when the most recent

 

 

user-initiated Sync or Idle change has propagated through to

all modules in a system. The change is a result of asserting Sync or forcing Idle via the Control Register or issuing a meas_control command or function