Image 128

id

pio1

pio2

Note

dir

nrdy

Agilent E1439 Programmer's Reference

Functions listed alphabetically

age1439_fiber_xmt_signals

Sets the transmitted values of PIO1, PIO2, DIR, and NRDY FPDP control signals on the fiber transmitter.

VXIplug&play Syntax

#include "age1439".h

ViStatus age1439_fiber_xmt_signals(ViSession id, ViInt16 pio1, ViInt16 pio2, ViInt16 dir, ViInt16 nrdy);

Description

These are embedded Serial FPDP signals. The use of these bits is optional. Serial FPDP does not use these four signals directly, but simply transmits them from sender to receiver. These functions set the value of PIO1, PIO2, DIR and NRDY bits on the fiber transmitter. The default value of these signals is 0.

Parameter

is the VXI instrument session pointer returned by the age1439_init function.

Programmable I/O bit on the fiber transmitter for user defined purposes.

Programmable I/O bit on the fiber transmitter for user defined purposes.

AGE1439_FIBER_SIG_ON FPDP control signals enabled.

AGE1439_FIBER_SIG_OFF FPDP control signals disabled. This is the default value for all signals.

The following are FPDP signals that are accommodated in the Serial FPDP protocol. For further information on these signals refer to ANSI/VITA 17-1998, Front Panel Data Port Specifications.

sets the dir FPDP control signal.

sets the nrdy FPDP control signal.

Return Value

AGE1439_SUCCESS indicates that a function was successful.

Values other than AGE1439_SUCCESS indicate an error condition or other important status condition. To determine the error message, pass the return value to “age1439_error_message” on page 102.