RATE CALIBRATION

RATE CALIBRATION is probably not necessary. The software initially expects STANDARD hardware. However, if you are using an auger feeder on a model with four dispense valves, or a 1/2 inch auger instead of a 1" auger, the software will take 10 or 20 cycles to adjust to correct operation. This slows down the process for a short time, but does not effect accuracy.

If your hardware is STANDARD, or

 

you prefer letting the system SELF ADJUST,

 

you may:

 

 

+---------------------------------------------------

+

¦

 

¦

¦ PROCEED TO:

NORMAL OPERATION

NEXT PAGE ¦

¦

 

¦

+---------------------------------------------------

+

To perform a RATE CALIBRATION:

Have all HOPPERS LOADED with enough material to run for several cycles without running out.

Have the "OFF AT END OF CYCLE" and "PAUSE" switches DOWN.

Press:

*

Display

will

say:

(PASSWORD)

Press:

22222

Display

will

say:

(P

x)

If your system has auger feeders, operate them briefly to ensure they are fully primed. To do so:

Press:

OPER

Display will say:

(OPERATE )

Press:

COL

Run

until auger is dispensing.

Press:

ADD

Run

until

auger

is dispensing.

Press:

DUMP

This will

empty

the

weigh bin.

Now you can CALIBRATE EACH material.

To do so:

Press:

CAL

Display

will

say:

(CALIBRATE)

Press:

REG

REGRIND

will

calibrate itself.

For each material, press CAL followed by the proper component key. Each time a dispense will occur, followed by weighing, followed by a dump to empty the weigh bin.

Press:

EXIT

when all calibrations are done.

 

 

YOU ARE NOW READY FOR ACCURATE PRODUCTION BLENDING.

 

 

+---------------------------------------------------

 

 

+

 

¦

 

 

 

¦

 

¦ PROCEED TO:

SPECIAL INSTRUCTIONS

NEXT PAGE

¦

 

¦

 

 

 

¦

 

+---------------------------------------------------

 

 

+

Page 16