MODEL 54eA

SECTION 5.0

 

SOFTWARE CONFIGURATION

TABLE 5-1. Program Settings List

(continued)

 

 

ITEM

 

 

CHOICES

FACTORY SETTINGS

CONFIGURE

 

 

 

 

A. Display options (Section 5.5)

 

 

 

1.

Measurement

Oxygen, ozone, free chlorine, total chlorine, monochloramine

Oxygen

2.

Sensor (Oxygen only)

 

Rosemount standard, Rosemount biopharm, or

Rosemount standard

 

 

 

 

other steam sterilizable

 

3.

Units (Oxygen only)

 

 

ppm, ppb, % saturation

ppm

4. Temperature units

 

 

°C or °F

°C

5.

Output 1

 

 

mA or % of full scale

mA

6.

Output 2

 

 

mA or % of full scale

mA

7.

Language

 

English, Français, Español, Deutsch, Italiano

English

8.

Main display left

 

 

See section 5.5

Sensor current

9.

Main display right

 

 

See section 5.5

Output 1 current

10 Display contrast

 

 

00-99 (darkest)

50

11. Test timeout

 

 

On or off

On

12. Timeout value

 

 

1 to 60 min

10 min

B Outputs (Section 5.6)

 

 

 

 

1.

Output 1 Control

 

 

 

 

 

a. Measurement

 

Oxygen, chlorine, ozone, pH, or temperature

Oxygen

 

b. Control

 

 

Normal or PID

Normal

2.

Output 1 Setup (normal)

 

 

 

 

 

a. Current

 

 

4-20 mA or 0-20 mA*

4-20 mA

 

b. Dampening

 

 

0-299 sec

0 sec

 

c. Hold mode

 

 

Hold last value or go to fixed value

Hold last value

 

d. Fixed hold value

 

 

0-22 mA

21 mA

 

e. Fault value

 

 

0-22 mA

22 mA

3.

Output 1 Setup (PID)

 

 

 

 

 

a. Proportional

 

 

0 to 299.9%

100 %

 

b. Integral

 

 

0 to 2999 sec

0 sec

 

c. Derivative

 

 

0 to 299.9%

0/0%

4.

Output 2 Control

 

 

 

 

 

a. Measurement

 

Oxygen, chlorine, ozone, pH, or temperature

Temperature

 

b. Control

 

 

Normal or PID

Normal

5.

Output 2 Setup (normal)

 

 

See output 1

See output 1

6.

Output 2 Setup (PID)

 

 

See output 1

See output 1

7.

Hold feature

 

 

Enable, disable, or 20 min timeout

Disable

C. Alarms (Section 5.7)

 

 

 

 

1.

Alarm 1 Control

 

 

 

 

 

a. Activation method

 

Oxygen, chlorine, ozone, temperature, pH

Oxygen

 

b. Control mode

 

 

Normal or TPC

Normal

2.

Alarm 1 setup (normal)

 

 

 

 

 

a. Configuration

 

 

Low, high, or off

High

 

b. Hysteresis

 

 

 

 

 

if oxygen (ppm)

 

 

0 to 20 ppm

0 ppm

 

if oxygen (ppb)

 

 

0 to 999 ppb

0 ppb

 

if oxygen (% saturation)

 

 

0 to 200%

0%

 

if chlorine, monochloramine, or ozone

-9999 to 9999 ppm

0 ppm

 

if pH

 

 

0 to 14.00

0 pH

 

if temperature

 

 

0 to 10°C

0.1°C

 

c. Delay time

 

 

0-99 sec

0 sec

 

d. Relay fault

 

 

none, open, closed

None

 

 

 

 

 

 

*Option-09, HART-enabled version operates at 4-20 mA only on output 1.

18

Continued on the following page

 

Page 24
Image 24
Emerson Process Management 53eA instruction manual Choices Factory Settings Configure