MODEL 54eA

SECTION 5.0

 

SOFTWARE CONFIGURATION

5.10NOISE REDUCTION

1.For maximum noise reduction the frequency of the ac power must be entered into the analyzer.

Temperature Noise rejection Main sensor cal

2.Press any key to enter the main menu. Move the cursor to "Program" and press Enter (F4).

3.Move the cursor to "Configure" and press Enter (F4).

4.Move the cursor to "Noise rejection" and press Enter (F4).

ExitEnter

 

 

5. To change the frequency setting, press Edit (F4). Use the é key to

Noise rejection

: 60 Hz

toggle between 50 and 60 Hz. Press Save (F4) to store the change.

ExitEdit

42

Page 48
Image 48
Emerson Process Management 53eA instruction manual 60 Hz, Noise Reduction