MODEL 54eA | SECTION 12.0 |
| CALIBRATION - OZONE |
SECTION 12.0
CALIBRATION - OZONE
12.1 INTRODUCTION
As Figure
The zero standard is necessary because ozone sensors, even when no ozone is in the sample, generate a small current called the residual current. The analyzer compensates for the residual current by subtracting it from the measured current before converting the result to an ozone value. New sensors require zeroing before being placed in service, and sensors should be zeroed whenever the electrolyte solution is replaced. Either of the fol- lowing makes a good zero standard:
•Deionized water.
•Tap water known to contain no ozone. Expose tap water to
The purpose of the
•Take the grab sample from a point as close to the sensor as possible. Be sure that taking the sample does not alter the flow of the sample to the sensor. It is best to install the sample tap just downstream from the sensor.
•Ozone solutions are unstable. Run the test immediately after taking the sample. Try to calibrate the sensor when the ozone concentration is at the upper end of the normal operating range.
FIGURE 12-1. Sensor Current as a Function of Ozone Concentration
70