| MODEL 54eA | SECTION 5.0 | 
| 
 | SOFTWARE CONFIGURATION | 
TABLE 5-1.  Program Settings List (continued)
| 
 | ITEM | CHOICES | FACTORY SETTINGS | 
| CONFIGURE (continued) | 
 | 
 | |
| D. pH measurement (Section 5.8) | 
 | 
 | |
| 1. pH sensor | Enable or disable | Enable | |
| 2. pH compensation (free chlorine only) | Auto or manual | Auto | |
| 3. pH value | 0.00 - 14.00 | 7.00 | |
| 4. pH diagnostic | On or off | Off | |
| 
 | a. Glass impedance high | 1000 MΩ | |
| 
 | b. Glass impedance low | 20 MΩ | |
| 
 | c. Reference impedance high | 140 kΩ | |
| 
 | d. Zero offset | 60 mV | |
| 
 | e. Impedance temp. compensation | On or off | On | 
| 5. pH calibration | 
 | 
 | |
| 
 | a. Auto calibration (buffer list) | Manual, Standard, Merck, Ingold, DIN 19267 | Standard | 
| 
 | b. Stabilize pH | 0.01 - 0.50 | 0.1 | 
| 
 | c. Stabilization time | 10 sec | |
| 6. | Solution temperature coefficient | 
 | 
 | 
| 
 | a. Temperature coefficient | 0.000 | |
| 
 | b. Operating isopotential | 7.00 | |
| 
 | c. Sensor isopotential | 0 to 14 pH | 7.00 | 
| E. Temperature compensation (Section 5.9) | 
 | ||
| 1. Temperature compensation | Auto or manual | Auto | |
| 2. | Manual temperature | 25°C | |
| F. Noise Reduction (section 5.10) | 
 | 
 | |
| Noise rejection | 50 or 60 Hz | 60 Hz | |
| G. Main sensor calibration (Section 5.11) | 
 | ||
| 1. | Stabilize reading | 
 | 
 | 
| 
 | a. oxygen (ppm) | 0 to 20 ppm | 0.05 ppm | 
| 
 | b. oxygen (ppb) | 0 to 999 ppb | 50 ppb | 
| 
 | c. oxygen (% saturation) | 0 to 200% | 1% | 
| 
 | d. chlorine, monochloramine, or ozone | 0 to 9999 ppm | 0.05 ppm | 
| 2. | Stabilize time | 0 - 30 sec | 10 sec | 
| 3. | Sensor zero stabilization value | 
 | 
 | 
| 4. | Dual range calibration | Enable or disable | disable | 
| 5. | Salinity (parts per thousand) | 0.0 - 99.9 o/oo | 0.0 o/oo | 
| H. Barometric Pressure (Section 5.12) | 
 | 
 | |
| 1. | Barometric pressure (oxygen only) | Auto or manual | Auto | 
| 2. | Barometric pressure manual | 0.4 - 1.2 atm (or equivalent units) | 1.00 atm | 
| 3. | Barometric pressure units | mm Hg, in Hg, bar, kPa | mm Hg | 
| I. Security (Section 5.13) | 
 | 
 | |
| 1. | Lock all | 000 | |
| 2. | Lock program | 000 | |
| 3. | Lock configuration | 000 | |
| 
 | 
 | 
 | 
 | 
20