MODEL 54eA |
| SECTION 16.0 | ||
|
|
| TROUBLESHOOTING | |
16.10 TROUBLESHOOTING NOT RELATED TO MEASUREMENT PROBLEMS | ||||
|
|
|
| |
| Problem | Action |
| |
| Display segments missing | Replace display board |
| |
| Alarm relays are chattering | 1. | Check alarm setpoints. | |
|
| 2. Increase hysteresis time delay settings (see Section 5.7) |
| |
| Incorrect current output | 1. | Verify that output load is less than 600 Ω. | |
|
| 2. | For minor errors, trim outputs (see Section 13.0) | |
|
| 3. | Replace power supply board |
|
| Display too light or too dark | Change contrast (see Section 5.5) |
| |
| “Level 1, 2 or 3 security: Lock” shown in display | Controller has password protection (see Section 5.13) |
| |
| “Hold mode activated” showing in display | Controller is in hold (see Section 5.6, steps 10 and 11) |
| |
| “Simulating output 1 or 2” showing in display | Controller is simulating outputs (see Section 5.4) |
| |
| “Simulating alarm 1, 2, 3 or 4” showing in display | Controller is simulating alarms (see Section 5.4) | ||
|
|
|
|
|
16.11 SIMULATING INPUTS - DISSOLVED OXYGEN
To check the performance of the controller, use a decade box to simulate the current from the oxygen sensor.
A.Disconnect the anode and cathode leads from terminals 1 and 2 on TB3 and connect a decade box as shown in Figure
B.Set the decade box to the resistance shown in the table.
| Sensor | Polarizing Voltage | Resistance | Expected current |
|
| 499ADO | 34 kΩ | 20 μA |
| |
| 499A TrDO | 20 kΩ | 40 μA | ||
|
|
|
|
|
|
| Hx438 and Gx448 | 8.4 MΩ | 80 nA | ||
|
|
|
|
|
|
C.Note the sensor current. To view the sensor current from the main display, press any key to enter the main menu. Move the
cursor to "Diagnostics" and press Enter (F4). The sensor cur- rent is the second line in the display. Note the units: μA is microamps, nA is nanoamps.
D.Change the decade box resistance and verify that the correct current is shown. Calculate current from the equation:
voltage (mV)
current (μA) =
resistance (kΩ)