MODEL 54eA | SECTION 5.0 |
| SOFTWARE CONFIGURATION |
SECTION 5.0
SOFTWARE CONFIGURATION
The instrument is configured at the factory to measure oxygen.
Figure 5-1 is an outline of the menu structure.
Table 5-1 lists the default settings and the range of choices available for each setting. To reduce the chance of error when configuring the controller the first time, enter settings in the order shown in the table.
TABLE 5-1. Program Settings List
ITEM | CHOICES | FACTORY SETTINGS |
SETPOINTS |
|
|
A. Alarms (Section 5.1) |
|
|
1. Alarm 1 (low action) |
|
|
a. if oxygen (ppm) | 0 ppm | |
b. if oxygen (ppb) | 0 ppb | |
c. if oxygen (% saturation) | 0 to 200% | 0 % |
d. if chlorine, monochloramine, or ozone | 0 ppm | |
e. if pH | 0.00 | |
f. if temperature | 0.1°C | |
2. Alarm 2 (high action) |
|
|
a. if oxygen (ppm) | 20 ppm | |
b. if oxygen (ppb) | 900 ppb | |
c. if oxygen (% saturation) | 0 to 200% | 200 % |
d. if chlorine, monochloramine, or ozone | 20 ppm | |
e. if pH | 14.00 | |
f. if temperature | 130°C | |
3. Alarm 3 | See alarm 2 | See alarm 2 |
B. Outputs (Section 5.2 and 5.3) |
|
|
1. Output 1 or 2: 4 mA setting |
|
|
a. if oxygen (ppm) | 0 ppm | |
b. if oxygen (ppb) | 0 ppb | |
c. if oxygen (% saturation) | 0 to 200% | 0 % |
d. if chlorine, monochloramine, or ozone | 0 ppm | |
e. if pH | 0.0 | |
f. if temperature | 0.1°C | |
2. Output 1 or 2: 20 mA setting |
|
|
a. if oxygen (ppm) | 20 ppm | |
b. if oxygen (ppb) | 900 ppb | |
c. if oxygen (% saturation) | 0 to 200% | 200% |
d. if chlorine, monochloramine, or ozone | 20 ppm | |
e. if pH | 14.00 | |
f. if temperature | 130°C | |
3. Setpoint (PID) |
|
|
a. if oxygen (ppm) | 1.00 ppm | |
b. if oxygen (ppb) | 100 ppb | |
c. if oxygen (% saturation) | 0 to 200% | 100% |
d. if chlorine, monochloramine, or ozone | 1.00 ppm | |
e. if pH | 7.00 | |
f. if temperature | 25°C | |
|
|
|
Continued on the following page | 17 |
|