MODEL 54eA

SECTION 5.0

 

SOFTWARE CONFIGURATION

SECTION 5.0

SOFTWARE CONFIGURATION

The instrument is configured at the factory to measure oxygen.

Figure 5-1 is an outline of the menu structure.

Table 5-1 lists the default settings and the range of choices available for each setting. To reduce the chance of error when configuring the controller the first time, enter settings in the order shown in the table.

TABLE 5-1. Program Settings List

ITEM

CHOICES

FACTORY SETTINGS

SETPOINTS

 

 

A. Alarms (Section 5.1)

 

 

1. Alarm 1 (low action)

 

 

a. if oxygen (ppm)

-99 to 99 ppm

0 ppm

b. if oxygen (ppb)

-999 to 999 ppb

0 ppb

c. if oxygen (% saturation)

0 to 200%

0 %

d. if chlorine, monochloramine, or ozone

-9999 to 9999 ppm

0 ppm

e. if pH

-2.00 to 20.00

0.00

f. if temperature

-5 to 130°C

0.1°C

2. Alarm 2 (high action)

 

 

a. if oxygen (ppm)

-99 to 99 ppm

20 ppm

b. if oxygen (ppb)

-999 to 999 ppb

900 ppb

c. if oxygen (% saturation)

0 to 200%

200 %

d. if chlorine, monochloramine, or ozone

-9999 to 9999 ppm

20 ppm

e. if pH

-2.00 to 20.00

14.00

f. if temperature

-5 to 130°C

130°C

3. Alarm 3

See alarm 2

See alarm 2

B. Outputs (Section 5.2 and 5.3)

 

 

1. Output 1 or 2: 4 mA setting

 

 

a. if oxygen (ppm)

-99 to 99 ppm

0 ppm

b. if oxygen (ppb)

-999 to 999 ppb

0 ppb

c. if oxygen (% saturation)

0 to 200%

0 %

d. if chlorine, monochloramine, or ozone

-9999 to 9999 ppm

0 ppm

e. if pH

-2.00 to 20.00

0.0

f. if temperature

-5 to 130°C

0.1°C

2. Output 1 or 2: 20 mA setting

 

 

a. if oxygen (ppm)

-99 to 99 ppm

20 ppm

b. if oxygen (ppb)

-999 to 999 ppb

900 ppb

c. if oxygen (% saturation)

0 to 200%

200%

d. if chlorine, monochloramine, or ozone

-9999 to 9999 ppm

20 ppm

e. if pH

-2.00 to 20.00

14.00

f. if temperature

-5 to 130°C

130°C

3. Setpoint (PID)

 

 

a. if oxygen (ppm)

-99 to 99 ppm

1.00 ppm

b. if oxygen (ppb)

-999 to 999 ppb

100 ppb

c. if oxygen (% saturation)

0 to 200%

100%

d. if chlorine, monochloramine, or ozone

-9999 to 9999 ppm

1.00 ppm

e. if pH

-2.00 to 20.00

7.00

f. if temperature

-5 to 130°C

25°C

 

 

 

Continued on the following page

17

 

Page 23
Image 23