17

MODEL 54eA SECTION 5.0
SOFTWARE CONFIGURATION
SECTION 5.0SOFTWARE CONFIGURATION

The instrument is configured at the factory to measure oxygen.

Figure 5-1 is an outline of the menu structure.

Table 5-1 lists the default settings and the range of choices available for each setting. To reduce the chance of

error when configuring the controller the first time, enter settings in the order shown in the table.

TABLE5-1. Program Settings List

Continued on the following page
ITEM CHOICES FACTORY SETTINGS
SETPOINTS
A. Alarms (Section 5.1)
1. Alarm 1 (low action)
a. if oxygen (ppm) -99 to 99 ppm 0 ppm
b. if oxygen (ppb) -999 to 999 ppb 0 ppb
c. if oxygen (% saturation) 0 to 200% 0 %
d. if chlorine, monochloramine, or ozone -9999 to 9999 ppm 0 ppm
e. if pH -2.00 to 20.00 0.00
f. if temperature -5 to 130°C 0.1°C
2. Alarm 2 (high action)
a. if oxygen (ppm) -99 to 99 ppm 20 ppm
b. if oxygen (ppb) -999 to 999 ppb 900 ppb
c. if oxygen (% saturation) 0 to 200% 200 %
d. if chlorine, monochloramine, or ozone -9999 to 9999 ppm 20 ppm
e. if pH -2.00 to 20.00 14.00
f. if temperature -5 to 130°C 130°C
3. Alarm 3 See alarm 2 See alarm 2
B. Outputs (Section 5.2 and 5.3)
1. Output 1 or 2: 4 mA setting
a. if oxygen (ppm) -99 to 99 ppm 0 ppm
b. if oxygen (ppb) -999 to 999 ppb 0 ppb
c. if oxygen (% saturation) 0 to 200% 0 %
d. if chlorine, monochloramine, or ozone -9999 to 9999 ppm 0 ppm
e. if pH -2.00 to 20.00 0.0
f. if temperature -5 to 130°C 0.1°C
2. Output 1 or 2: 20 mA setting
a. if oxygen (ppm) -99 to 99 ppm 20 ppm
b. if oxygen (ppb) -999 to 999 ppb 900 ppb
c. if oxygen (% saturation) 0 to 200% 200%
d. if chlorine, monochloramine, or ozone -9999 to 9999 ppm 20 ppm
e. if pH -2.00 to 20.00 14.00
f. if temperature -5 to 130°C 130°C
3. Setpoint (PID)
a. if oxygen (ppm) -99 to 99 ppm 1.00 ppm
b. if oxygen (ppb) -999 to 999 ppb 100 ppb
c. if oxygen (% saturation) 0 to 200% 100%
d. if chlorine, monochloramine, or ozone -9999 to 9999 ppm 1.00 ppm
e. if pH -2.00 to 20.00 7.00
f. if temperature -5 to 130°C 25°C