MODEL 54eA

SECTION 5.0

 

SOFTWARE CONFIGURATION

5.7 CHANGING ALARM PARAMETERS (continued)

Activate : Process

ExitEdit

Activation method Control mode

c.To change the activation method, press Edit (F4). Use the é key to scroll through the choices: "Process", "Temperature", and “pH” (if pH was enabled). "Process" means the measurement made by the main sensor (oxygen, ozone, free chlorine, or total chlo- rine). Press Save (F4) to store the selection.

d.The display returns to the "Activate: Process" screen. Press Exit (F1). The display returns to the "Activation method" screen. Move the cursor to "Control mode" and press Enter (F4).

ExitEnter

Ctrl mode : Normal

e.To change the control mode, press Edit (F4). Use the é key to toggle between "Normal" and "TPC". Press Save (F4) to store the selection.

ExitEdit

Alarm 1 control

Alarm 1 setup

Alarm 2 control

ExitEnter

Alarm : Low

Setpoint: 0.000 ppm

Hysteresis: 0.000 ppm

ExitEdit

8.Alarm setup for normal alarms:

a. Move the cursor to the desired alarm setup and press Enter (F4).

b.Use the é and ê keys to move the cursor to the desired param- eter. Press Edit (F4). Use the arrow keys to change the setting to the desired value and press Save (F4) to store the value. See the Figures 5-2 and 5-3 for an explanation of terms: low alarm, high alarm, hysteresis, and delay. See Table 5-1 for allowed values and limits.

Relay default determines how the relay will operate if there is a fault or the controller is in hold. Alarms can be forced on (Close), off (Open), or remain unchanged (None).

34

Page 40
Image 40
Emerson Process Management 53eA instruction manual Activate Process, Alarm 1 control, Alarm 1 setup, Alarm 2 control