MODEL 54eA

SECTION 8.0

 

CALIBRATION - FREE CHLORINE (499ACL-01)

8.2ZEROING THE SENSOR

1.Place the sensor in the zero standard (see Section 8.1). Be sure no air bubbles are trapped against the mem-

brane. The sensor current will drop rapidly at first and then gradually reach a stable zero value. To monitor the sensor current, go to the main display. Press any key to obtain the main menu. Press the ê key once to high-

light “Diagnostic variables.” Press Enter (F4). The sensor current is the second item in the display. Note the units: nA is nanoamps, μA is microamps. To return to the main display, press exit (F1) twice. Typical zero cur- rent for a free chlorine sensor is -10 to +10 nA.

A new sensor or a sensor in which the electrolyte solution has been replaced may require several hours (occa- sionally as long as overnight) to reach a minimum zero current. DO NOT START THE ZERO ROUTINE UNTIL THE SENSOR HAS BEEN IN ZERO SOLUTION FOR AT LEAST TWO HOURS.

Calibrate main sensor

Zero main sensor Adjust temperature

ExitEnter

0.0 ppm

Sensor must be in zero solution

Abort Cont

2.From the main display, press any key to obtain the main menu. With the cursor on “Calibrate,” press Enter (F4).

NOTE

If Hold was enabled in Section 5.6, the hold screen will appear. To activate hold, refer to Section 5.6, step 11.

Press the ê key once to move the cursor to “Zero main sensor.” Press Enter (F4).

3.The screen at left appears. Press Cont (F3). “Wait” flashes until the sensor is stabilized. Once the zero step is complete, the message “Sensor zero done” appears.

4.Press Exit (F1) three times to return to the main display.

NOTE

If Hold was activated during calibration, “Hold Mode Activated” will continue to flash in the main display. Return the sensor to normal and deactivate Hold. Refer to Section 5.6, step 11.

NOTE

During calibration, ERROR and WARNING messages may appear. If an ERROR message appears, press Exit (F1) to leave and return to the previous screen. If a WARNING message appears, press Cont (F3) to continue the calibration or press Abort (F1) to leave. Continuing the calibration after a warning message appears may cause substantial errors in the subsequent measurement. Refer to Section 15.5 for assis- tance.

56

Page 62
Image 62
Emerson Process Management 53eA instruction manual Calibrate main sensor, Zeroing the Sensor