MODEL 54eA

SECTION 8.0

 

CALIBRATION - FREE CHLORINE (499ACL-01)

8.3FULL SCALE CALIBRATION

1.Place the sensor in the process liquid. If automatic pH correction is being used, calibrate the pH sensor (see Section 11.0) and place it in the process liquid. If manual pH correction is being used, measure the pH of the process liquid and enter the value. See Section 5.8. Adjust the sample flow until it is within the range recommended for the chlorine sensor. Refer to the sensor instruction sheet.

2.Adjust the chlorine concentration until it is near the upper end of the control range. Wait until the controller reading is sta- ble before starting the calibration.

Calibrate main sensor Zero main sensor Adjust temperature

ExitEnter

1.100ppm

Stabilizing . . . Wait

Abort Cont

1.100ppm

Calibrate : 1.100 ppm

ExitEdit

3.From the main display, press any key to obtain the main menu. With the cursor on “Calibrate,” press Enter (F4).

NOTE

If Hold was enabled in Section 5.6, the hold screen will appear. To acti- vate hold, refer to Section 5.6, step 11.

Press Enter (F4).

4.Press Cont (F3). “Wait” flashes until the sensor is stabilized.

If the controller appears locked, the reading is not stable enough. Wait until the process readings are stable before starting the calibration.

Alternatively, increase the stability concentration or reduce the stability time. See Section 5.11. Calibrating while readings are unstable may sub- stantially reduce accuracy.

5.Once the reading is stable, the screen at left appears. Sample the process liquid. Make a note of the reading before taking the sample. Immediately determine free chlorine. Note the controller reading again. If the present reading (X) differs from the reading when the sample was taken (Y), cal- culate the value to enter (C) from the following formula:

C = (X/Y) (A)

where A is the concentration of chlorine measured in the grab sample.

Press Edit (F4). Use the arrow keys to change the concentration in the sec- ond line of the display to the desired value. Press Save (F4) to store the value.

6.Press Exit (F1) four times to return to the main display.

NOTE

If Hold was activated during calibration, “Hold Mode Activated” will continue to flash in the main display. Return the sensor to normal and deactivate Hold. Refer to Section 5.6, step 11.

7.During the calibration, the analyzer stores the measured current and cal- culates the sensitivity. Sensitivity is the sensor current in nA divided by the measured concentration. The sensitivity of the 499ACL-01 (free chlorine) sensor is 250-350 nA/ppm at 25°C and pH 7.

To view the sensitivity from the main display, press any key to enter the main menu. Press the ê key once. Then press Enter (F4) to display the

diagnostic variables. The sensitivity is the third line on the screen. Note the units: nA is nanoamps, μA is microamps.

NOTE

During calibration, ERROR and WARNING messages may appear. If an ERROR message appears, press Exit (F1) to leave and return to the previous screen. If a WARNING message appears, press Cont (F3) to continue the cali- bration or press Abort (F1) to leave. Continuing the calibration after a warning message appears may cause substan- tial errors in the subsequent measurement. Refer to Section 15.5 for assistance.

57

Page 63
Image 63
Emerson Process Management 53eA instruction manual Calibrate 1.100 ppm, Full Scale Calibration