Emerson Process Management 53eA instruction manual Section, PH at 25C Standards Nominal pH

Models: 53eA

1 116
Download 116 pages 57.39 Kb
Page 79
Image 79

MODEL 54eA

SECTION 13.0

 

CALIBRATION - pH

SECTION 13.0

CALIBRATION - pH

13.1 INTRODUCTION

A new pH sensor must be calibrated before use. Regular recalibration is also necessary.

A pH measurement cell (pH sensor and the solution to be measured) can be pictured as a battery with an extreme- ly high internal resistance. The voltage of the battery depends on the pH of the solution. The pH meter, which is basically a voltmeter with a very high input impedance, measures the cell voltage and calculates pH using a con- version factor. The actual value of the voltage-to-pH conversion factor depends on the sensitivity of the pH sens- ing element (and the temperature). The sensing element is a thin, glass membrane at the end of the sensor. As the glass membrane ages, the sensitivity drops. Regular recalibration corrects for the loss of sensitivity. pH cali- bration standards, also called buffers, are readily available.

Two-point calibration is standard. Both automatic calibration and manual calibration are available. Auto calibration avoids common pitfalls and reduces errors. Its use is recommended.

In automatic calibration the controller recognizes the buffer and uses temperature-corrected pH values in the cal- ibration. The table below lists the standard buffers the controller recognizes. The controller also recognizes sever- al technical buffers: Merck, Ingold, and DIN 19267. Temperature-pH data stored in the controller are valid between at least 0 and 60°C.

 

pH at 25°C

 

Standard(s)

 

 

 

Note 1: NIST is National Institute of Standards,

 

(nominal pH)

 

 

 

 

 

DIN is Deutsche Institute für Normung, JSI is

 

1.68

 

NIST, DIN 19266, JSI 8802, BSI (see note 1)

 

Japan Standards Institute, and BSI is British

 

Standards Institute.

 

3.56

 

NIST, BSI

 

 

 

 

 

 

 

Note 2: pH 7 buffer is not a standard buffer. It is

 

3.78

 

NIST

 

 

 

 

 

 

 

a popular commercial buffer in the United

 

4.01

 

NIST, DIN 19266, JSI 8802, BSI

 

 

 

States.

 

 

 

 

 

 

 

6.86

 

NIST, DIN 19266, JSI 8802, BSI

 

 

 

 

 

 

7.00

 

(see note 2)

 

 

 

 

 

 

7.41

 

NIST

 

 

 

 

 

 

9.18

 

NIST, DIN 19266, JSI 8802, BSI

 

 

 

 

 

 

10.01

 

NIST, JSI 8802, BSI

 

 

 

 

 

12.45

 

NIST, DIN 19266

 

 

 

 

 

During automatic calibration, the controller also measures

 

 

 

 

 

 

 

 

 

 

noise and drift and does not accept calibration data until

 

 

 

 

 

readings are stable. Calibration data will be accepted as soon

 

 

 

 

 

as the pH reading is constant to within the factory-set limits of

 

 

 

 

 

0.02 pH units for 10 seconds. The stability settings can be

 

 

 

 

 

changed. See Section 5.11.

 

 

 

 

 

In manual calibration, the controller still monitors readings for

 

 

 

 

 

stability; however, the buffer pH lookup feature is missing.

 

 

 

 

 

The user has to enter the correct pH at the temperature the

 

 

 

 

 

buffer is being used.

 

 

 

 

 

 

Once the controller completes the calibration, it calculates

 

 

 

 

 

the calibration slope and offset. The slope is reported as the

 

 

 

 

 

slope at 25°C. Figure 13-1 defines the terms.

 

 

 

 

 

The controller can also be standardized. Standardization is

 

 

 

 

 

 

FIGURE 13-1. Calibration Slope and Offset

 

the process of forcing the controller reading to match the

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

reading from a second pH instrument. Standardization is

 

 

 

 

 

sometimes called a one-point calibration.

 

 

 

 

 

73

Page 79
Image 79
Emerson Process Management 53eA instruction manual Section, PH at 25C Standards Nominal pH