MODEL 54eA

SECTION 7.0

 

CALIBRATION - DISSOLVED OXYGEN

SECTION 7.0

CALIBRATION - DISSOLVED OXYGEN

7.1 INTRODUCTION

As Figure 7-1 shows, oxygen sensors generate a current directly proportional to the concentration of dissolved oxygen in the sample. Calibrating the sensor requires exposing it to a solution containing no oxygen (zero stan- dard) and to a solution containing a known amount of oxygen (full-scale standard).

The zero standard is necessary because oxygen sensors, even when no oxygen is present in the sample, gener- ate a small current called the residual current. The analyzer compensates for the residual current by subtracting it from the measured current before converting the result to a dissolved oxygen value. New sensors require zeroing before being placed in service, and sensors should be zeroed whenever the electrolyte solution is replaced. The recommended zero standard is 5% sodium sulfite in water, although oxygen-free nitrogen can also be used.

The Model 499A TrDO sensor, used for the determination of trace (ppb) oxygen levels, has very low resid- ual current and does not normally require zeroing. The residual current in the 499A TrDO sensor is equivalent to less than 0.5 ppb oxygen.

The purpose of the full-scale standard is to establish the slope of the calibration curve. Because the solubility of atmospheric oxygen in water as a function of temperature and barometric pressure is well known, the natural choice for a full-scale standard is air-saturated water. However, air-saturated water is difficult to prepare and use, so the universal practice is to use air for calibration. From the point of view of the oxygen sensor, air and air-sat- urated water are identical. The equivalence comes about because the sensor really measures the chemical poten- tial of oxygen. Chemical potential is the force that causes oxygen molecules to diffuse from the sample into the sensor where they can be measured. It is also the force that causes oxygen molecules in air to dissolve in water and to continue to dissolve until the water is saturated with oxygen. Once the water is saturated, the chemical potential of oxygen in the two phases (air and water) is the same.

Oxygen sensors generate a current directly proportional to the rate at which oxygen molecules diffuse through a membrane stretched over the end of the sensor. The diffusion rate depends on the difference in chemical poten- tial between oxygen in the sensor and oxygen in the sample. An electrochemical reaction, which destroys any oxy- gen molecules entering the sensor, keeps the concentration (and the chemical potential) of oxygen inside the sen- sor equal to zero. Therefore, the chemical potential of oxygen in the sample alone determines the diffusion rate and the sensor current.

When the sensor is calibrated, the chemical potential of oxygen in the standard determines the sensor current. Whether the sensor is calibrated in air or air-saturated water is immaterial. The chemical potential of oxygen is the same in either phase. Normally, to make the calculation of solubility in common units (like ppm DO) simpler, it is convenient to use water-saturated air for calibration.

Automatic air calibration is standard. The user simply exposes the sensor to water-saturated air. The controller monitors the sensor current. When the current is stable, the controller stores the current and measures the baro- metric pressure and temperature. The temperature element is part of the dissolved oxygen sensor. The pressure sensor is inside theanalyzer. From the temperature, the

controller calculates the saturation vapor pressure of water. Next, it calculates the pressure of dry air by sub- tracting the vapor pressure from the barometric pressure. Using the fact that dry air always contains 20.95% oxy- gen, the analyzer calculates the partial pressure of oxy- gen. Once the analyzer knows the partial pressure of oxy- gen, it uses the Bunsen coefficient to calculate the equi- librium solubility of atmospheric oxygen in water at the prevailing temperature. At 25°C and 760 mm Hg, the equilibrium solubility is 8.24 ppm.

Often it is too difficult or messy to remove the sensor from

 

 

 

the process liquid for calibration. In this case, the sensor

 

 

 

can be calibrated against a measurement made with a

 

 

 

portable laboratory instrument. The laboratory instrument

 

 

 

 

FIGURE 7-1. Sensor Current as a Function of

 

typically uses a membrane-covered amperometric sensor

 

 

 

Dissolved Oxygen Concentration

 

that has been calibrated against water-saturated air.

 

 

 

 

 

 

 

 

49

Page 55
Image 55
Emerson Process Management 53eA instruction manual Section Calibration Dissolved Oxygen, Dissolved Oxygen Concentration