MODEL 54eA

SECTION 5.0

 

SOFTWARE CONFIGURATION

5.7 CHANGING ALARM PARAMETERS

1.This section describes how to configure the controller alarms. Alarms 1, 2, and 3 can be assigned to the main amperometric measurement (oxygen, ozone, free chlorine, or total chlorine), temperature, or pH (if pH was enabled). In addition, alarm 1, 2, or 3 can be configured as a feed limit timer or as an interval timer (see steps 10 and 11). Alarm 4 is always a fault alarm.

2.An alarms assigned to the main amperometric measurement, temperature, or pH can be configured as either a simple on/off alarm or as TPC (time proportional control) alarm.

Output setpoints

Simulated tests

Configure

ExitEnter

3.Press any key to enter the main menu. Move the cursor to "Program" and press Enter (F4).

4.Move the cursor to "Configure" and press Enter (F4).

Outputs Alarms pH

5. Move the cursor to "Alarms" and press Enter (F4).

Exit

Enter

 

 

6. Nine menu headers relate to alarms. Alarms 1, 2 and 3, each have a

Alarm 1 control

 

 

control header and a setup header. Alarm 4 has only a setup header.

Alarm 1 setup

The eighth menu header is for configuring the feed limit timer, and the

ninth menu header is for configuring the interval timer.

 

 

Alarm 2 control

Always configure the control parameters BEFORE making

 

 

changes in the alarm setup.

Exit

Enter

 

 

To access a header, move the cursor to the desired header and press

 

 

Enter (F4).

 

 

7. Alarm Control Settings:

Activation method

a. Move the cursor to the desired output control header. Press Enter

Control mode

(F4).

 

 

b. With the cursor on "Activation method" press Enter (F4).

Exit

Enter

33

Page 39
Image 39
Emerson Process Management 53eA Changing Alarm Parameters, Always configure the control parameters Before making