MODEL 54eA

SECTION 7.0

 

CALIBRATION - DISSOLVED OXYGEN

7.2ZEROING THE SENSOR

1.Place the sensor in a fresh solution of 5% sodium sulfite (Na2SO3) in water. Be sure air bubbles are not trapped against the membrane. The current will drop rapidly at first and then gradually reach a stable zero

value. To monitor the sensor current, go to the main display. Press any key to obtain the main menu. Press

the ê key once to highlight “Diagnostic variables.” Press Enter (F4). The sensor current is the second item in the display. Note the units: nA is nanoamps, μA is microamps. To return to the main display, press exit (F1) twice.The table gives typical zero currents for Rosemount Analytical sensors.

 

Sensor

Zero current (nA)

 

 

499ADO

<50 nA

 

 

499A TrDO

<5 nA

 

 

Hx438 and Gx448

<1 nA

 

 

 

 

The Model 499A TrDO does not normally require zeroing.

A new sensor or a sensor in which the electrolyte solution has been replaced may require several hours (occa- sionally as long as overnight) to reach a minimum current. DO NOT START THE ZERO ROUTINE UNTIL THE SENSOR HAS BEEN IN ZERO SOLUTION FOR AT LEAST TWO HOURS.

Calibrate main sensor Zero main sensor Adjust temperature

ExitEnter

2.From the main display, press any key to obtain the main menu. With the cursor on “Calibrate,” press Enter (F4).

NOTE

If Hold was enabled in Section 5.6, the hold screen will appear. To activate hold, refer to Section 5.6, step 11.

3.Press the ê key once to move the cursor to “Zero main sensor.” Press Enter (F4).

0.0 ppm

Sensor must be in zero solution

Abort Cont

4.The screen at left appears. Press Cont (F3). “Wait” flashes until the sensor is stabilized. Once the zero step is complete, the message “Sensor zero done” appears.

5.Press Exit (F1) three times to return to the main display.

NOTE

If Hold was activated during calibration, “Hold Mode Activated” will continue to flash in the main display. Return the sensor to normal and deactivate Hold. Refer to Section 5.6, step 11.

NOTE

During calibration, ERROR and WARNING messages may appear. If an ERROR message appears, press Exit (F1) to leave and return to the previous screen. If a WARNING message appears, press Cont (F3) to continue the calibration or press Abort (F1) to leave. Continuing the calibration after a warning message appears may cause substantial errors in the subsequent measurement. Refer to Section 15.4 for assis- tance.

50

Page 56
Image 56
Emerson Process Management 53eA instruction manual Sensor must be in zero solution, Zeroing the Sensor