Advanced Operations

3.Press Edit.

An automatic menu guidance queries values (filters, wavelength, nominal values and tolerances) given in the quality control certificate, to the following specifications:

Stray Light

Photometrical accuracy

Wavelength accuracy

4.Press OK when all values are entered and the overview is displayed.

5.Press Verification.

6.Insert the Multi Cell Holder in the cell compartment in such a way that the cell opening for 10 mm square cuvettes/cells is at the front. Secure the holder with two locking screws.

7.Remove any cuvettes/cells from the cell compartment, close the cell compartment, and press Start.