Standard Operations

5.2 Instrument Setup mode

1.Select Instrument Setup in the Main Menu.

A selection of functions appears in order to configure the functions of the instrument.

This display appears when the multi-cell holder is installed or the 13 mm round cuvette/vial compartment is used. If the imstrument recognizes a Sipper module or a Peltier module the Carousel Options key is replaced by Sipper Options or Peltier Options.

Optional modules for the DR5000:

Carousel Holder (Sample Changer)

Pour-Thru Cell module

Sipper module

Peltier module (Temperature control module)

Refer to the corresponding manuals/instruction sheets for more information.

5.2.1 Operator ID

Use this option to enter up to 30 sets of operator initials (up to

5 characters each) into the instrument. This feature helps record which operator measured each sample.

1.Press Operator ID in the Instrument Setup.

2.Press New to enter a new Operator ID.

3.Use the alphanumeric keypad to enter a new Operator ID.

4.Press OK to confirm.

5.The display shows the chosen Operator ID.

6.Press OK. The instrument will return to the Instrument Setup screen and show the selected operator identifier.

7.The chosen Operator ID is activated.

Note: Press Delete to remove an Operator ID from the list.

Note: Alternatively, enter or change an Operator ID in measurement mode. In the results screen, press Options>More>Instrument Setup or if an Operator ID is already assigned, select the "Operator ID symbol immediately in the results screen.

23

Page 23
Image 23
Hach DR 5000 user manual Instrument Setup mode, Standard Operations, Operator ID