Standard Operations

 

Table 5 Barcode program options (continued)

Options

Description

 

 

Send Data icon /

To send data to a printer, computer or USB memory stick (Type A)

Send Data

 

 

 

 

This functions as a stopwatch. It helps to ensure that the steps of an analysis are correctly timed

Timer icon

(e.g. reaction times, wait times, etc., can be exactly specified). When the specified time has

elapsed, an acoustic signal is emitted. The use of the timer has no influence on the measurement

 

 

program.

 

 

 

A corrective dilution factor can be entered in order to take account of certain properties.

 

The number entered at the dilution factor prompt will be multiplied by the result to compensate for

Dilution Factor Off/On

the adjustment. For example, if the sample has been diluted by a factor of 2, enter 2. The default

setting of the dilution factor is turned off.

 

 

Note: When a dilution is in effect, the dilution icon will appear on the display.

 

Note: If undiluted samples are used, set the dilution factor off.

Chemical Form

Some of the stored tests/methods allow to select the chemical form and the measuring range.

 

 

Edit

To modify an existing program

 

 

Save as User Program

To store the selected parameters as a User Program, see section 6.1 on page 61.

 

 

Recall Data

Call up saved measurement data, wavelength scans or time courses, see section 5.3 on

page 35.

 

Instrument Setup

Basic data of the instrument, see section 5.2 on page 23.

 

 

5.5.5 Sample blank

Turbidity and color in the sample matrix can falsify the results of a photometric analysis. The interference factors come from the sample or are created by reactions with the reagents.

The influence of turbidity and/or color can be eliminated or reduced by taking a sample blank reading.

In the barcode mode, a special cuvette/vial (TNT919) containing the sample blank is placed in cell compartment #1 after the sample reading has been taken and is automatically measured. The sample reading is then corrected by adding or subtracting the blank value. The final result is displayed, with the message "After blank value corr.".

Some barcode tests do not require a sample blank value to be determined, as turbidity and color are dealt with during the test

53

Page 53
Image 53
Hach DR 5000 user manual Sample blank, Options Description Send Data icon, Edit