Hach DR 5000 user manual Wavelength scan reading, Scale & units Press Scale & Units

Models: DR 5000

1 136
Download 136 pages 38.22 Kb
Page 97
Image 97

Advanced Operations

4.Press Return to return to the scan mode.

Note: The Integral is shown instead of the date on the display.

Note: For the next scan measurement the setting for the Integral will be

On.

Scale & units

1.Press Scale & Units.

2.Highlight the required units (Abs or %T).

3.Highlight Auto or Manual scaling on the graph´s y-axis.

Note: If manual scaling is selected, use the alphanumeric keypad to set the limits ymin. and ymax.. The graph is adjusted to display only the values in the selected range. If automatic scaling is selected, the instrument sets the limits automatically so that the total range can be displayed.

4.Press OK.

5.Press Return to return to the scan mode.

6.6.2 Wavelength scan reading

After the scanning parameters have been selected, the baseline must be scanned. Changing any of the scanning parameters requires a new baseline scan. When the baseline has been scanned, the instrument is ready to scan one or more samples.

1.Press Wavelength Scan in the Main Menu.

2.Insert the blank cuvette/cell into the cuvette/sample cell holder.

97

Page 97
Image 97
Hach DR 5000 user manual Wavelength scan reading, Scale & units Press Scale & Units