Hach DR 5000 user manual For data storage, see .3.1.1 on

Models: DR 5000

1 136
Download 136 pages 38.22 Kb
Page 93
Image 93

Advanced Operations

6.5.2Complete a measurement in the multi wavelength mode

1.Insert the blank cuvette/cell into the cuvette/sample cell holder. Press Zero.

Note: The Read key does not become active until the zero measurement has been completed.

2.Insert the sample cuvette/cell into the cuvette/sample cell holder. Press Read.

Note: The UV lamp is not switched on until a wavelength in the UV spectrum has been entered and the blank reading has been started. During the warm-up phase of the UV lamp, the message ”Warming up...“ is displayed and the ”UV lamp“ symbol flashes. As soon as the UV lamp is ready, the blank reading is carried out.

Note: Zero and Read are disabled until the cell compartment is closed.

3.For data storage, see section 5.3.1.1 on page 35.

93