Hach DR 5000 user manual Verification kit, Press Nominal Values 113

Models: DR 5000

1 136
Download 136 pages 38.22 Kb
Page 113
Image 113

Advanced Operations

5.The Drift Check runs 1 hour.

This test takes a reading every minute for one hour. Every 15 minutes, linear regression is used to calculate the slope (rate of change) for the previous 15 minute interval.

6.The last reading and slope for each 15 minute interval are displayed. At the end of the hour, the overall values are calculated and displayed. The test proceeds to completion.

7.Press Cancel to return to Optical Checks.

6.8.3.6 Verification kit

The Verification Kit, LZV 537, (see Section 9 on page 129) is designed for periodic monitoring of scattered light, photometric accuracy and the wavelength accuracy of the spectrophotometers.

When results exceed allowable tolerances (given in the quality control certificate), contact the manufacturer.

1.Press Verification Kit in the Optical Checks menu.

2.Press Nominal Values.

113

Page 113
Image 113
Hach DR 5000 user manual Verification kit, Press Nominal Values 113