Advanced Operations

5.According to the selected wavelength range, select the interval from the actively displayed wavelength steps. In summery maximally 780 measuring steps can be accomplished during a Scan.

Note: The maximum wavelength adjusts automatically if the difference between the maximum and minimum wavelength is not a multiple of the interval.

6.Press OK to return to the scan mode. Selected parameters are displayed along the graph‘s x-axis.

Select view (displaying table)

1.Press Select View in the Options menu after a reading is taken.

2.A table with the results is displayed.

3.To return to the graph press Options and then View Graph.

Cursor mode

1.Press Cursor Mode: Track in the Options menu.

2.The selection for this menu item determines what data are displayed in the table. Highlight Track or Peak/Valley.

3.Press OK to confirm.

4.Press Return to return to the scan mode.

Integral

The Integral applies to the whole wavelength range of the scan.

1.Press Integral: Off in the Options menu.

2.Highlight On to show the Integral. To find the integral of other wavelength ranges, change the wavelength range and scan again.

3.Press OK to confirm.

96

Page 96
Image 96
Hach DR 5000 user manual Select view displaying table, Cursor mode, Integral