Advanced Operations

 

Table 13 Multi-wavelength setup options (continued)

Options

Description

 

 

Recall Data

Call up saved measurement data, wavelength scans or time courses, see section 5.3 on

page 35.

 

 

 

Instrument Setup

Basic data of the instrument, see section 5.2 on page 23.

 

 

λ/ Absorbance formula:

1.Press Absorbance Formula.

2.The formula selected in the top key determines the number of wavelength and coefficent keys that will appear below. To change the absorbance formula, press the top key, select a formula from the displayed list and press OK. When a new formula is selected, the number of variables below changes to match.

The following formulas are available:

K1 A 1 + K 2 A 2

K 1 A 1 + K 2 A 2 + K 3 A 3

K1 A 1 + K 2 A 2 + K 3 A 3 + K 4 A 4 K 1 A 1 / K 2 A 2

(K 1 A 1 + K 2 A 2 ) / K 3 A 3

(K 1 A 1 + K 2 A 2 ) / (K 3 A 3 + K 4 A 4 )

A 1 refers to the absorbance at wavelength 1

A2 refers to the absorbance at wavelength 2, etc. K 1 refers to the coefficient at wavelength 1

K 2 refers to the coefficient at wavelength 2, etc.

Coefficients can be set negative where subtraction is required.

91