Section 1

Specifications

 

 

Specifications are subject to change without notice.

 

 

Performance specifications

 

 

 

 

Operating mode

 

Transmittance (%), Absorbance and Concentration

 

 

 

Source lamp

 

Gas-filled Tungsten (visible) and Deuterium (UV)

 

 

 

Wavelength range

 

190–1100 nm

 

 

 

Wavelength accuracy

 

± 1 nm in Wavelength Range 200–900 nm

 

 

Wavelength reproducibility

< 0.1 nm

 

 

 

Wavelength resolution

 

0.1 nm

 

 

 

Wavelength calibration

 

Automatic

 

 

 

Wavelength selection

 

Automatic, based on method selection

 

 

 

Scanning speed

 

900 nm/min 1 nm steps

 

 

 

Spectral bandwidth

 

2 nm

 

 

 

Photometric range

 

± 3.0 Abs in Wavelength Range 200–900 nm

 

 

 

Photometric accuracy

 

5 mAbs at 0.0–0.5 Abs

 

1% at 0.50–2.0 Abs

 

 

 

 

 

 

 

< 0.5%–2 Abs

Photometric linearity

 

< = 1% at > 2 Abs

 

 

with neutral glass at 546 nm

 

 

 

Stray light

 

KI-solution at 220 nm

 

> 3.3 Abs / < 0.05%

 

 

 

 

 

Data storage

 

1000 measured values (result, date, time, sample ID, user ID)

 

 

 

User programs

 

50

 

 

Physical and environmental specifications

 

 

 

Width

 

450 mm (17.7 in)

 

 

 

Height

 

200 mm (7.9 in)

 

 

 

Depth

 

500 mm (19.7 in)

 

 

 

Weight

 

15.5 kg (34.2 Ibs)

 

 

 

Operating requirements

 

10–40 °C (50–104 °F), max. 80% relative humidity (non-condensing)

 

 

 

Storage requirements

 

–25–60 °C (–13–140 °F) max. 80% relative humidity (non-condensing)

 

 

Additional technical data

 

 

 

 

Mains connection

 

100–240 V~; 50/60 Hz; automatic changeover

 

 

 

 

 

Use only shielded cable with max. length of 3 m.

Interfaces

 

1 x USB type B

 

 

2 x USB type A

 

 

 

Water-resistant to

 

IP 31

Enclosure

 

 

 

 

 

 

Safety class

 

Safety class I

 

 

 

7

Page 7
Image 7
Hach DR 5000 user manual Section Specifications