4.8.2 Battery Replacement

WARNING

The [] illumination indication shows lithium battery deterioration, which is used for memory backup. The battery has a lifetime of approximately 10 years under normal operation.

When [] is illuminated, contact Omega Engineering.

Replacing the lithium battery can be dangerous. Do not attempt to replace the battery; never disassemble the recorder for replacing the battery.

4.8.3 Fuse Replacement !

Replacing the fuse every two years is recommended as preventive maintenance.

Replacing procedure is as follows:

(1)Turn the power switch OFF.

(2)Take out the chart cassette; the fuse holder is found in the main unit. Turn the fuse carrier counterclockwise to release it with the fuse.

(3)Change the fuse with a new one, and insert the carrier to the fuse holder turning it clockwise for fixing.

WARNING

Fuse Replacement

To protect against personal injury, turn the power switch OFF and disconnect the recorder from the main power line before replacing the fuse.

To protect against an accident such as a fire, use only the specified fuse

Fuse part number:

A1360EF

 

Original manufacturer:

Shelter

Littel

Original manufacturer’s product code:

FST0034.524

313.500

Ratings:

250V AC T 0.5 A

250VACT 0.5 A

Recorders with a gray power switch

 

 

Fuse part number:

A1512EF

 

Original manufactturer:

wickmann-Werke GmbH

Original manufacturers product code:

No.19195

 

Ratings:

250V AC T0.8A

 

Recorders with a white power switch

 

 

4.8.4 Cleaning the Carriage Shafts