6.1 Setting Input Range and Recording Span (RANGE)

Difference Computation Setting (DELT Setting)

To obtain the difference between the measured values of two channels, proceed as follows.

Key Operation

ESC

DISP

3 sec

PRINT

CH UP

PRINT

CH UP

PRINT

CH UP

PRINT

CH UP

Indication

Description

Press the [DISP] key for three seconds to enter the SET mode.

Select the [ ] display by using the [ ] key. Then press the [ENT] key.

Use the [ ] key to select the desired channel.

Note that channel No.1 cannot be of the ‘DELT’ type. Press the [ENT] key.

Use the [ ] key to select [

].

Press the [ENT] key.

 

The display which appears will allow you to enter a reference channel number. Note that the reference channel number must be lower than that of the channel being set. Therefore channel

No.1 cannot be of the [ ] type.

Use the [ ] key to select the desired reference channel number. Press the [ENT] key.

PRINT FEED

CH UP

PRINT FEED

CH UP

 

ESC

 

CH UP

DISP

3 sec

 

 

In case of VOLT the display which appears will show the same minimum value of the recording span as of the reference channel.

Use the [ ] and [ ] keys to select the minimum value (SPAN L).

Press the [ENT] key.

Note that in case of VOLT, the input range of the DELT-channel must be within the range of the reference channel. In case of TC, RTD, the minimum/maximum span values can be found by taking the ±value of the input range width. E.g. input range of TC type L = –200 to 900°C: minimum/maximum span values will be at ±1100 °C. The difference in measured value from the reference channel will be recorded on the DELT-channel.

The display which appears will show the same maximum value of the recording span as of the reference channel.

Use the [ ] and [ ] keys to select the maximum value (SPAN R) in the same way as described under ‘SPAN L setting’. Note that SPAN L cannot be equal to SPAN R. Press the [ENT] key.

[ ] is displayed and the setting is completed. Press the [ENT] key to set another channel; or press the [DISP] key for three seconds to return to the OPERATION mode.

6. SET MODE

NOTE If you press the [ESC] key in the middle of setting, the indication returns to the