Calibration is a comparing operation between standard inputs and measured values. It is recommended to carry out a calibration once a year to assure the measuring accuracy.

Required instruments for calibration

The instruments for calibration should have the required resolution. If necessary, ask the sales office from whom you purchased the recorder about those instruments.

Recommended instruments

OMEGA model # CL511 calibrator

Calibration procedure

(1)Connect the instruments to the recorder; connecting the terminals is shown below. Give enough warm-up time to the instruments (more than 30 minutes for the recorder).

(2)Check that the ambient temperature and humidity are within the standard operating

conditions.

Standard operating conditions

Ambient temperature :

23±2 ˚C

Humidity

:

55±10% RH

Power supply

:

90 to 132 V AC, or 180 to 250 V AC

Power frequency

:

50/60 Hz±0.1%

Warm-up time

:

30 minutes or more

Mechanical vibration : Within the range not affecting recorder operation

However, when the calibration is conducted under normal operating conditions (see subsection 4.8.6), check the accuracy influenced by the conditions.

Chart paper yields thermal expansion or shrinkage, sometimes resulting in differences from the calibrated values.

Examples of expansion are shown below assuming the reference temperature 20˚C and humidity 65% RH: Expansion at 85% RH ..... 0.4% or less

Expansion at 35% RH ..... 0.45% or less

(3)Input the standard signals equivalent to 0, 50, and 100% of the setting range of the recorder, then check the difference error between input signals and measured values.

If the error is not within the specified accuracy, contact OMEGA Engineering.

NOTE A reference input for a thermocouple (TC) signal needs temperature compensation; the temperature at the recorder input terminal should be measured to make voltage compensation (adding) due to the reference junction temperature compensation.

Power terminals

1

2

L

N

+

3

+

DC voltage standard

Input terminals

(for DC Voltage or TC)

DC Voltage Signal for Pen model (same as Dot-printing model)

Temperature Measurement for Resistance Temperature Detector (RTD)

4 - 14

Page 39
Image 39
Omega RD260A instruction manual Required instruments for calibration, Recommended instruments, Calibration procedure