6.5 Setting Chart Speed (CHART)

6.5 Setting Chart Speed (CHART)

Chart Speed (unit: mm/h)

10

15

20

25

30

40

50

60

75

80

 

 

 

 

 

 

 

 

 

 

90

100

120

150

160

180

200

240

300

360

375

450

600

720

750

900

1200

1500

1800

2400

 

 

 

 

 

 

 

 

 

 

3000

3600

4500

4800

5400

6000

7200

9000

10800

12000

 

 

 

 

 

 

 

 

 

 

The data for a dot-printing model (10 mm/h to 1500 mm/h) are shown within the bold border.

Key operation

ESC

DISP

3 sec

PRINT

CH UP

PRINT

CH UP

Indication

Description

Press the [DISP] key for three seconds to enter the SET mode,

and select [ ] using the [ ] key. Then press the [ENT] key.

Using the [ ] key, select one of the two kinds of chart speed, the No.1 and No.2 chart speeds. Then press the [ENT] key.

NOTE

PRINT

CH UP

NOTE

 

ESC

 

CH UP

DISP

3 sec

 

 

The No.2 chart speed identifies the chart speed which takes effect in case the speed is changed by remote control (optional function).

For models without the remote control function, the setting of the No.2 chart speed is invalid.

Select a numeric code referring to the table above using the [ ] key and press the [ENT] key.

Digital printing will not be carried out when the chart speeds are: 120 mm/h or more for a dot-printing model

When a pen model uses a fast chart speed such as 600 mm/h to 1500 mm/h, the position error of the time tick of a periodic printout may become as large as 2 cm for 1500 mm/h.

[ ] is dispalyed and the setting is completed. Press the [ENT] key to set another channel; or press the [DISP] key for three seconds to return to the OPERATION mode.

6. SET

NOTE If you press the [ESC] key in the middle of setting, the indication returns to the

[ ] and already set data will be aborted.

MODE

6 - 15

Page 66
Image 66
Omega RD260A instruction manual Setting Chart Speed Chart, Chart Speed unit mm/h