7.10 Turning Alarm Printing On/Off

7.10 Turning Alarm Printing On/Off

When alarms occur, the channel number, level, type of alarm and ON/OFF time can be printed on the right side of the chart. It is selectable to get the ON/OFF time of alarms (ON1), the ON time only (ON2) or to get no printout (OFF). Note that alarms will not be printed when the chart speed exceeds 1800 mm/h (Pen model) or 120 mm/h (Dot- printing model).

The initial value is ‘ON1’, and this setting is common for all channels.

SETTING PROCEDURE:

Enter the SET UP mode by turning ‘ON’ the power while pressing and holding the [CH

UP] (ENT) key until the SET UP display appears.

Use the [

] key to select the display [

]. Press the [ENT] key.

Use the [

] key to select the display [

]. Press the [ENT] key.

Select ‘ON1’ (ON/OFF time of alarms will be printed), ‘ON2’ (ON time only will be printed) or ‘OFF’ (no alarm will be printed). Press the [ENT] key.

The display [

] appears.

Before leaving the SET UP mode, you have to store your new settings. Use the [ ] key

to select the [ ] display, and press the [ENT] key.

Select [ ] to keep your new settings or [ ] to abort the settings, and press the [ENT] key. After a few seconds, the OPERATION mode will appear.