Siemens LU01 Quick Start P001 to P007, P001 Operation, P002 Material, P003 Maximum Process Speed

Models: LU01

1 114
Download 114 pages 47.65 Kb
Page 35
Image 35

Quick Start (P001 to P007)

P001 Operation

Sets the type of measurement required for the application.

 

0

 

Out-of-service

 

 

 

 

 

1

 

Level – how full the vessel is (Point 1 and/or 2 only)

Values

 

 

 

2

 

Space – how empty the vessel is (Point 1 and/or 2 only)

 

 

 

 

 

3

*

Distance – distance from transducer to material (Point 1 and/

 

or 2 only)

 

 

 

 

 

 

 

P002 Material

Specifies material type. If the material presents a flat surface perpendicular to the transducer beam, select liquid.

Values

1

*

Liquid or flat surface

2

 

Solid or angled surface

 

 

Alters

P830 TVT Type

P003 Maximum Process Speed

Determines level change reaction speed.

 

1

 

 

 

Slow (0.1 m/min)

 

 

 

 

 

 

 

2

 

*

 

Medium (1 m/min)

Values

 

 

 

 

 

3

 

 

 

Fast (10 m/min)

 

 

 

 

 

 

 

4

 

 

 

Surge (1.7 m/sec)

 

 

 

 

 

 

 

5

 

 

 

Immediate (17 m/sec)

 

 

P070

Failsafe Timer

 

P700 Max Fill Rate

 

P701 Max Empty Rate

 

P702

Filling Indicator

Alters

P703

Emptying Indicator

P704

Rate Filter

 

 

P710

Fuzz Filter

 

P713 Echo Lock Window

 

P727 Scan Delay

 

P841 Long Shot Number

 

Failsafe (P070 to P072)

 

Rate (P700 to P708)

Related

Measurement Verification (P710 to P713)

 

Transducer Scanning (P726 to P729)

 

P905

Transmit Pulse

 

 

 

 

 

 

Use a setting just fast enough to keep up with your process. Slower settings provide higher accuracy. Faster settings allow for more level fluctuations.

Parameters

7ML19985BE02

SITRANS LU01 – INSTRUCTION MANUAL

Page 29

Page 35
Image 35