Siemens LU01 P833 TVT Start Min, P834 TVT Start Duration, P835 TVT Slope Min, Range -30 to

Models: LU01

1 114
Download 114 pages 47.65 Kb
Page 83
Image 83

6.Press ARROWS to change the bias value of the breakpoint, –50 to 50.

7.Press ENTER .

8. Press TRANSDUCER

to update the scope display with the new TVT.

P833 TVT Start Min

Use this feature to adjust the TVT Curve height to ignore false echoes (or pick up true echoes) near the start of the Echo Profile.

Values

Range: -30 to 225

 

 

Preset: 50

 

 

 

 

Related

P800 Near Blanking

P834 TVT Start Duration

 

 

 

 

Enter the minimum TVT Curve start point (in dB above 1 μV rms).

This feature should only be used if increased Near Blanking (P800) would extend farther than desired into the measurement range.

P834 TVT Start Duration

Use this feature in conjunction with TVT Start Min (P833) to ignore false echoes (or pick up true echoes) near the start of the Echo Profile.

Values

Range: 0 to 9999

 

 

Preset: 30

 

 

 

 

Related

P833 TVT Start Min

P835 TVT Slope Min

 

 

 

 

Enter the time (in ms) for the TVT Curve to decrease from the TVT Start Min (P833) point to the TVT Curve baseline.

P835 TVT Slope Min

Enter the minimum slope (in dB/s) for the middle of the TVT Curve.

Values

Range: 0 to 9999

 

 

Preset: 200

 

Related

P830 TVT Type

P834 TVT Start Duration

 

 

 

 

Use this feature to adjust the slope declination, and use it in conjunction with TVT Start Duration (when a long flat TVT Type is selected) to ensure the TVT Curve remains above the false echoes in the middle of the Echo Profile. Alternatively, if TVT Type is set for TVT Slopes (P830 = 6), preset is 2000. Use this parameter to adjust the slope declination when TVT type is set to TVT slopes (P830=6).

Parameters

7ML19985BE02

SITRANS LU01 – INSTRUCTION MANUAL

Page 77