P001 Operation

Alters

P051 Maximum Volume

 

P920 Reading Measurement

P051 Maximum Volume

For Readings in volumetric units (rather than percent), enter the vessel volume between Empty (P006) and Span (P007).

Values

Range: 0.000 to 9999

 

 

Preset: 100.0

 

Alters

P060 Decimal Position

 

P006 Empty

Related

P007 Span

 

P924 Volume Measurement

 

 

 

Any volume units can be chosen because volume is calculated from empty to maximum span and is scaled according to the Tank Shape (P050) value.

Note: Make sure selected chosen units allow LCD volume display.

Examples:

If max. volume = 3650 m3, enter 3650

If max. volume = 267500 gallons, enter 267.5 (thousands of gallons)