TDS5000B Series Specificationsand Performance Verification 1-11
Table1- 2: Channel input and vertical specifications (Cont.)
Characteristic Description
DCvol tage measurement accuracy Measurement type DCaccuracy (in volt s)
Sample acquisition mode,
typical
Anysam ple ±[(1.5% + 1.0% ×|net
offset /offset range|)×
|reading --net offset| +
offset accuracy+ (0.13 di v
×V/div setting) + 0.6 mV]
TDS5104B: 2 mV/div --
3.98 mV/div
±[(1.5% + 3.0% ×|net
offset /offset range|)×
|reading --net offset| +
offset accuracy+ 0. 13 div
×V/div setting + 0.6 mV]
Delta voltage measurement
between any two points
acquiredunder the sam e
setup and ambient condi-
tions
±[1.5%+ 1.0% ×|net
offset /offset range|×
|reading --net offset| +
(0.26 div ×V/div setting) +
1.2 mV]
TDS5104B: 2 mV/div --
3.98 mV/div
±[1.5%+ 3.0% ×|net
offset /offset range|×
|reading --net offset| + 0.26
div ×V/div setting
+1.2mV]
where,net of fset = offset --(position ×volts/divisi on)
nAverageacquisiti on
mode
Averageof 16 waveforms ±[1.5%+ 1.0% ×|net
offset /offset range|×
|reading --net offset| +
offset accuracy+ 0. 06 div
×V/div]
TDS5104B: 2 mV/div --
3.98 mV/div
±[1.5%+ 3.0% ×
|net offset /offset range|×
|reading --net offset| +
offset accuracy+ 0. 06 div
×V/div]