Performance Tests
2-54 TDS5000B Series Specifications and PerformanceVerification
Output Signal Checks
The following proceduresverify characteristics of the output signals that are
listed as checkedunder Warranted Characteristics in the Specifications section.
The oscilloscopeoutputs these signals at its front and rear panels.
Equipment
required
Twoprecision 50 coaxial cables (Item 4)
Onecal ibration generator (Item 11)
Prerequisites ReadPrerequisites on page 2 --17and footnote warnings on page 2--19.
Also,the oscilloscope must have passedCheck DC Voltage
Measurement Accuracy on page 2--24.
1. Install the test hookupand preset the instrument controls:
TDS5000BSeries oscilloscope
50coaxial cables
Calibration
generator
To AUX OU T
(locatedon the back
ofthe instrument)
Figure2 -22: Initial test hookup
a. Hook up test-signal source1:
HConnect the standard amplitudeoutput of a calibrationgenerator
through a 50 precision coaxialcable to CH 3. See Figure 2--22.
HSet the calibration generatorto output a 0.500 V square wave.
b. Hook up test-signal source2: Connect the Aux Ou t at the rear panel to
CH 2 through a 50 precision cable.
c. Initialize the oscilloscope:Press the DEFAULT SETUP button.
d. Modify the initialized front-panelcontrol settings:
HPress the Vertical CH1button to toggle it off.
HPress the Vertical CH3button to display that channel.

Check Outputs:

CH 3 Signal Out

(TDS5034B, TDS5054B,

TDS5054BE TDS5104B)

and Aux TriggerOut