Performance Tests
2-44 TDS5000B Series Specifications and PerformanceVerification
HCalculate the difference of the Maximum (M) mi nus the mean ().
HCalculate the difference of the mean () minus the Minimum (m).
HBoth differencesmust be less than or equal to the Delta-time
accuracy limit shown in Table2--4 for your oscilloscope.
HEnter the pass/failresult for delta time in the test record.
Table2- 4: Delta time measurement
InstrumentModel Pulserise time range
Deltatim e accuracy
limit
TDS5104B, TDS5054B,
TDS5034B, TDS5052B,
TDS5032B
280ps --800 ps 0.060 ns
TDS5054BE 1.4 ns --4 ns (The internal
150M Hz bandwidth filter
will slow pulse rising edges
to this range.)
0.300 ns
e. Repeat for all other channels:
HNote the verticalscale setting of the channel just confirmed.
HPress the Verticalchannel button for the channel just confirmedto
remove thechannel from display.
HClick MEAS,andthenClearto remove the measurement.
HPress the front-panelbutton that corresponds to the channel that you
are to confirm.
HSet verticalS CALE to the settingnoted in step e, first bullet.
HPress the TriggerSource button to toggle the source to the channel
selected.
HMove the test hookup to the channel that you selected.
HFrom the toolbar,click the Vert button. Set the termination of the
channel to 50 by selectingthe channel tab and clicking the
Termination50 button. (If you are testing a TDS5054BE model,
make surethe 150 MHz bandwidth limit has been turned on. All
other models aretested at full bandwidth.)
HPressthe RUN/STOP button to start the display.
HRepeat step d.
2. Disconnect all testequipment from the oscilloscope.